Tinytag Talk 2 voltage/current wiring connections?

The wiring connections for our Tinytag Talk 2 voltage and current data loggers are as follows:

CAB-3233

For use with TK-4702 and TK-4802 data loggers.

Yellow: Signal input
Black: Common/0V
White: Sense line

The above information also applies to the now withdrawn TK-0702 Tinytalk voltage data logger.

Tinytag Plus Re-Ed voltage, current and count wiring connections?
The wiring connections for our Tinytag Plus Re-Educatable data loggers are as follows:

CAB-3239

For use with TGPR-0704, TGPR-0705, TGPR-0805 and TGPR-1002 data loggers.

Current Cable:

Yellow: Signal Input
Black: Common/0V
Red: 2.5V Reference
Green: 12V (NOT TGPR-0704)
White: Sense Line (TGPR-0704 ONLY)
OR:

Discontinued cable:

White: Signal input
Brown: Common/0V
Red: 2.5V reference
Black: 12V (NOT TGPR-0704)
Blue: Sense Line (TGPR-0704 ONLY)
CAB-3246

For use with TGPR-0804, TGPR-1001 and TGPR-1201 data loggers.

Current Cable:

Red: Signal input
Blue: Common/0V
OR:

Discontinued Cable:

Black: Signal input
Red: Common/0V