Tinytag Radio data loggers can still show low battery warnings
after their batteries have been replaced because their battery
voltage is only checked once every two hours.
To clear this warning, the data logger can be left until the
next battery check, after which the low battery warning will
disappear. Performing a Current Reading on a data
logger will also clear the warning.
If the data logger still shows a low battery warning after more
than two hours have passed, please contact us or
contact